Analysis of single U-rich particles by fission track laser ablation ICP-MS

Not scheduled
5m
M Building (Vienna International Centre)

M Building

Vienna International Centre

Wedge Participant Shaping the future of safeguards implementation (SGI) [SGI] Enhancements and Innovation in Sample Collection and Analysis

Speaker

Dr Stefanie Konegger-Kappel (IAEA Environmental Sample Laboratory)

Description

The isotopic analysis of single U-rich particles that are collected during IAEA inspections by means of cotton swipes is an important tool in nuclear safeguards to help verify the absence of undeclared nuclear material and activities. Two methods that are routinely used within the IAEA Network of Analytical Laboratories to determine particles’ U and/or Pu isotopic compositions are fission track thermal ionization mass spectrometry (FT-TIMS) and secondary ion mass spectrometry (SIMS). However, another method that has aroused interest as a complementary technique for particle analysis in the international safeguards community during the last decade is laser ablation coupled to inductively coupled plasma mass spectrometry (LA-ICP-MS).
U-rich particles that are collected during IAEA swipe sampling have sizes in the µm and sub-µm range and can typically not be identified with cameras mounted in laser ablation systems. In order to avoid determining “mixed” isotope ratios from two or more particles with different isotopic compositions, the location of the particles of interest (POI) needs to be identified prior to the LA-ICP-MS analysis. Here, a fission track technique, where natural U particles were deposited on Si wafers and irradiated at the Research Center Rez, Czech Republic, was used for selecting POIs for the subsequent LA-ICP-MS analysis. The fissions tracks are a good indicator for both enrichment and particle location. Coordinates were confirmed by scanning electron microscopy (SEM) analysis using a LYRA 3 FIB-SEM system (Tescan). The relocation accuracy between instruments was better than 20 µm.
The isotopic analysis of the selected particles was performed by LA – multi collector – ICP-MS, using a J200 fs-laser ablation system (Applied Spectra, Inc.) and a Neptune “Plus” MC-ICP-MS (Thermo Fisher Scientific). The determined 234U/238U and 235U/238U isotope ratios were overlapping within uncertainties with the certified values.
The presented work demonstrates very well the high potential of FT-LA-MC-ICP-MS for single particle analysis for environmental inspection samples.

Topics TEC5
Which "Key Question" does your Abstract address? TEC5.1
Which alternative "Key Question" does your Abstract address? (if any) TEC5.5

Primary author

Dr Stefanie Konegger-Kappel (IAEA Environmental Sample Laboratory)

Co-authors

Mr Felix Horak (IAEA Environmental Sample Laboratory) Dr Jan Lorincik (Research Center Rez) Ms Klara Rezankova (Research Center Rez) Ms Kristina Sihelska (Research Center Rez)